Part 2: Mastering Agarose Gel Electrophoresis: Key Factors for Visualizing and Quantifying DNA and RNA
By Gene Vantage – 4 September 2024
Agarose gel electrophoresis is an essential technique in molecular biology, widely used for the separation, visualization, and quantification of DNA and RNA. While the method is straightforward, achieving precise and reliable results requires careful attention to a number of critical variables. In this post, we’ll explore these key factors, providing practical tips to optimize your gel electrophoresis runs and strategies to address common challenges that may arise.
5. Using the Appropriate Ladder
A molecular weight ladder is essential for estimating the size of your DNA or RNA fragments by providing reference bands of known sizes. The ladder should cover the expected size range of your sample, with clear, distinct bands. Some ladders also contain loading dye, which can be convenient for tracking sample migration.
Always include a ladder in every gel run. Choose a ladder that spans the entire range of your expected fragment sizes. If you need to quantify DNA, it is helpful to select a ladder with bands of known mass per volume.
If your sample bands are not aligning with the ladder, ensure that you’re using the correct ladder for your gel’s concentration and running conditions. Verify that the ladder was not degraded or contaminated.
6. Proper Gel Setting and Preparation
For an agarose gel to function properly, it must set evenly and be free of bubbles. If the gel tray is disturbed during setting, the surface may become uneven, causing issues during sample loading and resulting in bands that run unevenly or merge.
Pour the gel gently to avoid bubbles, and allow it to set on a completely flat surface. Avoid moving the gel tray during the setting process to ensure the gel solidifies uniformly.
If bands are slanted or the sample wells are uneven, the gel may not have set correctly. Ensure that future gels are allowed to set undisturbed in a level position. If bubbles are present, pop them with a pipette tip immediately after pouring.
7. Complete Dissolution of Agarose Powder
Incomplete dissolution of agarose can result in an uneven gel matrix, leading to poor resolution and inconsistent migration. It is essential to heat the agarose-buffer mixture until the solution becomes completely clear, indicating full dissolution.
After heating, visually inspect the solution to ensure it is completely transparent and free of clumps or crystals. If using a microwave, heat in short bursts and swirl the flask between bursts to prevent local overheating.
If the gel appears cloudy or has irregularities, it might be due to incomplete dissolution. Reheat the solution until clear and pour the gel immediately after cooling slightly.
8. Accurate Buffer Preparation
The correct ratio of agarose to buffer is critical. Too much buffer results in a weak gel that may tear easily, while too little buffer leads to a brittle gel that can crack. Both situations can negatively impact the migration of DNA or RNA.
Use precision scales for weighing agarose and volumetric flasks or graduated cylinders for measuring buffers. Double-check your calculations, particularly when preparing solutions in large volumes.
If the gel does not set to the desired firmness or appears to tear or crack, review your agarose and buffer measurements. Adjust the ratio in subsequent preparations to achieve the desired gel consistency.
9. Choosing the Right Electrophoresis System
The type of electrophoresis system used can influence the outcomes of your gel runs. Horizontal systems are standard for DNA/RNA gels, while vertical systems are more often used for protein gels. However, factors such as buffer circulation, cooling, and the design of the gel chamber can impact the resolution and consistency of results.
Ensure your system has good buffer circulation and adequate cooling to prevent overheating. Regularly inspect the gel chamber for any damage or wear that might affect the run.
If you notice inconsistent results or unexpected heating, consider whether the system is functioning correctly. Check for issues with the power supply, buffer circulation, and gel placement.
10. Loading Samples and Dye Mobility
The loading buffer contains glycerol or another dense substance to help samples sink into the wells, and tracking dyes like bromophenol blue or xylene cyanol are added to monitor the progress of migration in the electrophoresis system. Different dyes migrate at different rates depending on the gel concentration and voltage applied.
Load your samples slowly to avoid spilling and ensure even loading across wells. Use a dye appropriate for your gel concentration and buffer system.
If the dye front migrates unexpectedly, verify the dye’s mobility characteristics under your specific running conditions. Adjust the concentration of the loading dye or select an alternative dye if necessary.
11. Choosing the Right Staining Agent
SYBR Green and Ethidium Bromide are commonly used stains for nucleic acids. While SYBR Green is often chosen for its high sensitivity and safety profile, Ethidium Bromide remains in use where cost, availability, or specific detection needs justify its continued application. In some cases, SYBR Green’s slightly different binding properties may influence the migration of very small DNA fragments, leading to subtle differences in band patterns compared to Ethidium Bromide. However, for most applications, SYBR Green provides a safer and equally effective alternative.
Choose a staining agent based on your lab’s safety protocols and sensitivity requirements. Ensure that the staining procedure is compatible with your downstream applications, such as sequencing or cloning, if gel extraction is required.
If bands are faint or if there’s high background noise, adjust the concentration of the stain. Also, consider switching to a more sensitive stain like SYBR Green if Ethidium Bromide is not providing adequate results.
Agarose gel electrophoresis is a highly reliable technique when all factors are carefully controlled. By understanding how each aspect—from agarose concentration to the choice of staining agent—impacts the outcome, you can achieve consistent and clear results. Always be vigilant for potential errors and be prepared to troubleshoot issues based on the symptoms observed. By applying these guidelines, you can master the art of nucleic acid separation, ensuring that your DNA and RNA visualization and quantification are both accurate and reproducible.